Product Description
Details
In case of muddy water, it suffices to add a small amount of distilled water (measured out in a graduated cylinder) to the gas generation tube.
Put the cap on the generation tube.
(By repeating this process several times, all sulphides become hydrogen sulphide several times repeatedly).
The scale in the detector tube is read, and the concentration of the sulphide in the sample is calculated.
Measuring Range |
0.002 to 0.02 mg |
Sampling Time |
45 seconds per pump stroke |
Detecting Limit |
0.0002 mg (n=1) |
Colour Change |
White Brown |
Reaction Principle |
Gus generation tube: Sulphides + H2SO4 H2S Detector tube: H2S + Pb(CH3COO)2 PbS |
Coefficient of Variation |
5% (for 0.002 to 0.02mg) |
Shelf Life |
3 Years |
Corrections for |
Unnecessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences |
|
Substance |
Concentration |
Interference |
Change colour by itself |
Chloride ion |
- |
No error |
No discoloration |
Chromate ion |
- |
No error |
No discoloration |
Carbonate ion |
- |
No error |
No discoloration |
Calibration gas generation |
Static gas dilution method |