Product Description
Details
GAS221L Gastec Chloride Ion Detector Tubes
The soluble ion tube system offers a quicker, cheaper and easier alternative to using laboratory based anaylsers for measuring dissolved substances
(i.e. in solution)
To measure, simply immerse the detector tube in a sample of liquid (placed in a beaker or other container). Anyone can measure concentrations quickly, easily and accurately on site.
Performance |
Measuring Range |
25 to 1000 mg/l |
Sampling Time |
3 minutes |
Detecting Limit |
1 mg/l |
Colour Change |
Brown to White |
Reaction Principle |
2Cl- + Ag2CrO4 (Brown) 2AgCl (White) |
Coefficient of Variation |
15% (for 25 to 300 mg/l), 10% (for 300 to 1000 mg/l) |
Shelf Life |
3 Years |
Corrections for |
Unnecessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences |
|
Substance |
Concentration |
Interference |
Change colour by itself |
CN- |
≥1 mg/l |
Plus error |
White discoloration |
Br- |
≥10 mg/l |
Plus error |
White discoloration |
SCN- |
≥30 mg/l |
Plus error |
White discoloration |
Fe2+ |
≥200 mg/l |
Minus error |
No discoloration (≤1000 mg/l) |
Fe3+ |
≥200 mg/l |
Minus error |
White discoloration |
F- |
≥500 mg/l |
Minus error |
No discoloration (≤1000 mg/l) |
I- |
≥50 mg/l |
Plus error |
White discoloration |
S2- |
≥5 mg/l |
Plus error |
Blackish grey discoloration |
SO42- |
≥2000mg/l |
Minus error |
White discoloration |
Calibration gas generation |
Sodium Chloride standard solution |