Product Description
Details
GAS221LL Gastec Chloride Ion Detector Tubes
The soluble ion tube system offers a quicker, cheaper and easier alternative to using laboratory based anaylsers for measuring dissolved substances
(i.e. in solution).
To measure, simply immerse the detector tube in a sample of liquid (placed in a beaker or other container). Anyone can measure concentrations quickly, easily and accurately on site.
Performance |
Measuring Range |
10 to 200 mg/l |
Sampling Time |
3 minutes |
Detecting Limit |
3 mg/l |
Colour Change |
Brown White |
Reaction Principle |
2Cl- + Ag2CrO4 (Brown) 2AgCl (White) + CrO42+ |
Coefficient of Variation |
15% (for 10 to 50 mg/l), 10% (for 50 to 200 mg/l) |
Shelf Life |
3 Years |
Corrections for |
Unnecessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences |
|
Substance |
Concentration |
Interference |
Change colour by itself |
CN- |
≥10 mg/l |
Plus error |
White discoloration (≥1 mg/l) |
Br- |
≥10 mg/l |
Plus error |
White discoloration (≥10 mg/l) |
SCN- |
≥1 mg/l |
Plus error |
White discoloration (≥100 mg/l) |
Fe2+ |
≥20 mg/l |
Minus error |
No discoloration (≤1000 mg/l) |
Fe3+ |
≥20 mg/l |
Plus error |
White discoloration (≥20 mg/l) |
F- |
≥1000 mg/l |
No error |
No discoloration (≤1000 mg/l) |
I- |
≥20 mg/l |
Plus error |
Pale yellow discoloration (≥20 mg/l) |
S2- |
≥3 mg/l |
Plus error |
Blackish grey discoloration (≥3 mg/l) |
SO42- |
≥1000mg/l |
No error |
No discoloration (≤1000 mg/l) |
Calibration gas generation |
Sodium Chloride standard solution |